  • Accuracy to within 1% for syringes >1ml
  • Precision of CV ≤0.05% at full stroke
  • Syringes sized between 50µl and 5ml

The MSP1-CXN is an OEM syringe pump designed for use inside lab automation systems and within IVD applications. It is highly accurate and precise with accuracy of greater than 1%. Syringes that fit inside the pump can be sized between 50µl and 5ml and are manufactured from borosilicate glass with a PTFE coated stainless steel plunger.

Multiple communication interfaces are available with RS485 and RS232 enabling control via an external computer or controller.

Useful 'How To' videos

How to operate the MSP1 CX Syringe Pump.

Firmware Update Utility of the MSP CX & SP CX Pump.

Other features

Main Spec

  • Accuracy:≤1.0 % (syringe ≥1 mL)
  • Precision:CV≤0.05% at full stroke (with syringe ≥ 1mL )
  • Wetted material:Borosilicate glass,PTFE,PFA,CTFE,ETFE,UHMW-PE
  • Dimension (L x W x H):110mm x 44.4mm x 127mm
  • Weight:1.0 kg
  • Power supply:DC 24V,1.5A (peak)

Syringe and Drive

  • Syringe fitting:1/4″-28 thread
  • Syringe:50uL, 100uL, 250uL, 500uL, 1mL, 2.5mL, 5mL
  • Syringe material:Borosilicate glass,stainless steel plunger with PTFE coating, PTFE or UHMW-PE plunger tip
  • Resolution (standard mode):3000 steps with 0.01mm in 1 step
  • Resolution (microstep mode):48000 microsteps with 0.625um in 1 microstep
  • Travel:30mm
  • Time for full stroke:1.26s-1200s
  • Drive system:Ball screw drive with optical encoder

Valve & Valve Drive

  • Turn Time:≤250ms between adjacent ports
  • Valve options:Standard: 3-port valve, optional: 3-port distribution valve, 4-port valve, T-valve, Y-block
  • Max pressure:0.68MPA
  • Fitting:1/4″-28 thread

Communication

  • Interface:RS485, RS232, CAN bus,external control input and output
  • Protocol:Data terminal,OEM protocol (serial) or CAN
  • Baud rate:"RS485 or RS232: 9600bps or 38400bps
  • CAN bus: 100 KB, 125 KB 250 KB, 500 KB or 1MB"
  • Format:Data bit: 8; parity: none; stop bit: 1; half duplex (RS232/485)

Firmware

  • Topology group:Up to 15 pumps in one group, support group control and individually addressed
  • Upgrade firmware online:"One Click" by customer through PC Utility
  • Homing algorithm:Plunger home position is identified by motor stall, which will reduce the dead volume of the system
  • Halt function:For better interactive control, "h " command is used to halt execution of the command string during plunger movement. To resume execution with "R" or "r" command
  • Microstep mode:Set microstep mode for smoother motion with 48000 microsteps per stroke, 1/8 microstepping or 1/16 microstepping
  • Programmable parameter:Programmable ramps, cut-off velocity, plunger speeds, backlash compensation, programmable work-flow command set , delays and loops, terminate moves, diagnostics, absolute and relative positions

Environmental

  • Operating temperature:Recommended: 15℃-40℃,use low temperature syringe for low temperature application
  • Operating humidity:<80% RH, no condensation
  • Storage temperature: -20℃-65℃
  • Storage humidity:<80% RH, no condensation